Development of some remote adaptable analytical methods and systems for analyses of radioactive samples
Loading...
Date
item.page.authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
A simple remote aliquoting system (RAS) and a remote titration
newlinesetup were developed for the remote analyses of highly radioactive
newlinesamples inside a hot cell. The remote calibration of the RAS by a
newlinetitrimetric method was validated and the systematic error and random error
newlineof the RAS were found to be less than 1%.
newlineA remote adaptable titration procedure was developed for the free
newlineacidity determination of solutions containing uranyl nitrate and plutonium
newlinenitrate without the addition of complexing agents. In this procedure, the
newlinetotal acid was first determined using a remote alkalimetric titration with a
newlinevisual indicator. The free acidity of the solution was calculated after
newlinesubtracting the contribution due to hydrolysable ions from the total
newlineacidity. The relative standard deviation and recovery of free acidity by this
newlinemethod were lt1% and 100 ± 5%, respectively, for aliquots containing
newlineuranium 0.5 15.9mg and plutonium 0.85 8.5mg. The nitric acid
newlineconcentration of the aliquots was in the range of 0.092 10.56 M. Since no
newline