Design and robust motion control of a planar hybrid manipulator for lower limb rehabilitation tasks
Loading...
Date
item.page.authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Rehabilitation robotics had proved its importance in the medical field,
newlinewhich had grown rapidly to deliver high dosage and intense physical training
newlinefor patients with a loco-motor disorder caused by stroke. As per the National
newlineSample Survey Organization (NSSO) survey of India, 27% of disabilities are
newlinedue to loco-motor. This in turn, increases the need for rehabilitation followed
newlineby rehabilitation robots. This research proposes a design and robust motion
newlinecontrol of the 1P-2PRP hybrid planar manipulator for lower limb
newlinerehabilitation treatments. The stationary system was designed for bed ridden
newlinepatients where they have slight or no control over their lower limbs to
newlineimprove the recovery in the early stages. At present, the commercial systems
newlineavailable were having poor dynamic performance and no supportive or
newlineorthosis systems. To overcome the above said issues hybrid planar
newlinemanipulator was proposed with a robust motion control scheme.
newlineThe proposed system was a hybrid planar manipulator with 1P-2PRP
newlineconfiguration as an active part of the robotic based rehabilitation system and
newlinethe orthosis with 3R configuration as passive orthosis support. The Hybrid
newlinemanipulator has fused advantages of both serial and parallel manipulators.
newlineThe therapeutic treatment of lower limb movements in the sagittal plane was
newlinefacilitated by the active part of the manipulator to induce therapeutic motions
newlineon the passive part (orthosis) that supports the patient as a Continuous Passive
newlineMotion (CPM) device.The effectiveness and performance of the proposed system along with the robustness of the control system has been demonstrated using a scaled down fabricated real-time prototype. Further the performance and sensitivityof the proposed control has been analyzed for different range of working conditions. The clinically obtained gait data set was provided to the system to perform a passive range of motion control and verified the feasibility of
newlineproposed rehabilitation robotic system. In addition, the clinical gait pattern
newlinegeneration and tracking task