Dosimetric studies on cyberknife robotic stereotactic radiosurgery
Loading...
Date
item.page.authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Radiotherapy is one of the major treatment methods for cancer
newlineThis field has undergone a greater extent of technological advancement in
newlinerecent decades One such innovation in the field of radiotherapy is the
newlineintroduction of industrial robot to guide the linear accelerators with high
newlineprecision and more degrees of freedom to aim the target This robot guided
newlineminiature linear accelerator is called as CyberKnife Accuray Inc Sunnyvale
newlineUSA The linear accelerator in the CyberKnife is a miniature flattening filter
newlinefree FFF accelerator Another important aspect in CyberKnife is the smaller
newlinefield sizes The maximum field size available in CyberKnife is 60mm
newlinediameter CyberKnife treatment fractionation regimens are hypofractionated
newlineCyberKnife is one among the few treatment units which are used to treat
newlineTrigeminal Neuralgia TN The smallest 5mm collimator is being used to
newlinetreat TN However the dosimetric accuracy of such small collimators is still
newlineunder research The Monte Carlo smoothing algorithms provided in the
newlineMultiplan CyberKnife treatment planning system Accuray Inc Sunnyvale
newlineUSA has five different dose smoothing algorithms The final smoothed dose
newlinedistribution of Monte Carlo calculations may differ based on the choice of
newlinesmoothing algorithm as the principle of smoothing is not the same Though
newlineCyberKnife is a technological innovation there are some unsorted features
newlineassociated with this
newlineIn the first study the dose linearity and MU stability characteristics
newlineof a CyberKnife stereotactic radiosurgery system were established All the 12
newlinefixedtype collimators starting from 5 to 60 mm were used for the dose
newlinelinearity study The dose linearity was examined in three MU small medium
newlineand large ranges The MU stability test was performed with 60 mm
newlinecollimator for 10 MU and 20 MU with different combinations The maximum
newlinedose linearity error of 388 was observed for 1 MU with 5 mm collimator
newlineDose linearity error in the small MU range was considerably higher than in
newlinethe medium and large MU ranges The maximum deviation in the MU
newlinestability wa